Advia 120 Hematology System
| Prepared by: | PM |
| Date Prepared: | March 20, 2001 |
| Code: | MP-006 |
| Reviewed by: | RVS |
| Revision #: | 3.0 |
| Date Revised: | 12/13/01 |
Calibrated commercial controls are run daily and must be within established ranges prior to analyzing samples.
| 0.0 | Abstract: The ADVIA 120 Hematology System is a diagnostic instrument that analyzes whole blood to provide the following types of results:
| ||||||||||||||||||||||
| 1.0 | Instrumentation Layout: n/a | ||||||||||||||||||||||
| 2.0 | Reagents and Expendables:
Bayer Diagnostics Division Bayer Corporation 511 Benedict Avenue Tarrytown, NY 10591-5097 Phone: (914)631-8000 Fax: (914)524-2132 http:\\www.bayerdiag.com Maintenance and upkeep provided by the Diagnostic Laboratory at TJL. | ||||||||||||||||||||||
| 3.0 | Setup: You need to contact the Diagnostic Laboratory to be sure that ADVIA is working and that controls have been run. Once ADVIA is operating properly, go and obtain the blood samples. | ||||||||||||||||||||||
| 4.0 | Run: Prepare Eppendorf tubes with 20 ul of 20% EDTA. Use hematocrit tubes coated with EDTA. Obtain 6 drops of blood in Eppendorf tubes. You want approximately 160 ul. or more of whole blood to run test. Stir sample with the HCT tube and flush out remaining blood into Eppendorf tube. Use a wooden toothpick to stir sample and to check for clots. Any clots in sample will cause it to fail the test. | ||||||||||||||||||||||
| 5.0 | Clean Up: Dispose of used Eppendorf tubes by placing them in a burn bin or hazardous waste container. Wipe up any blood spills with 70% ethanol. | ||||||||||||||||||||||
| 6.0 | Data Reduction: To save your specific file be sure to: Log On - Click New File - Save. At the end of the run, you can retrieve your data from ADVIA with a floppy disk by doing the following:
| ||||||||||||||||||||||
| 7.0 | Safety: Wear gloves when dealing with blood samples or controls. ADVIA TESTpoint Controls contain human blood cells in a preservative medium. | ||||||||||||||||||||||
| 8.0 | Time and Capacity: Calibration of controls takes about 25 minutes. Aside from the time it takes you to eyebleed, running samples takes about 1 minute per sample when done through open tube aspiration wand. | ||||||||||||||||||||||
| 9.0 | Protocols: Click Standby Button (located in left lower corner). Put control in incubator to warm. Start: Insert name and ID #; click Log On. Go To: Customize - system setup - systems options Go To: Run screen options - print none - save - routine operations Go to: Data options (left side of upper screen) Find: Raw data directory name (it should say C:\RAW_DATA/HLB). Click Save. You should see: System Options - Data 0 System Options Saved Click OK Go To: Routine Operations (left upper corner). Click Manual Sample ID. On this screen, click in the following information:
Before attempting to place sample, make sure top left corner says Ready to Run. Get controls and be sure to roll vial between hands to mix sample well. Two Ways to Place Sample in Advia:
1st: Primer has Pink Tape along the vial (pink tape will NOT print out any information - only used to wet the system). Go To: Manual Sample ID and click Sample Type to Control. Activate bar code on each tube with hand Opticon device. 2nd: Controls:
After each sample has run, you will see a lab screen pop up. Click Hide to go back to Manual Sample ID Screen. Then click QC at bottom of screen and review the lab results. Note: If you see Yellow, it could be that the sample was not mixed or warmed up enough. Re-do the sample. If you see Red, there is a problem with the sample and Sue Grindle (x6203) needs to be contacted. Go To: Routine Operations - Manual Sample ID - Change Sample Type to Primer 3rd: Primer Again (to wet system). Change sample type to Control. 4th: Controls:
Again click on Hide and return to Manual Sample ID Screen. Check lab values on QC screen. Observe for yellow or red flags. If all flags for controls are normal, proceed with eyebleed samples. For Eyebleed Samples: Go to Manual Sample ID, click New File - Save - OK. Make sure printer is set to print lab sheets by going to Customize - System set Up - System Options - Run Screen Options. Click Print All; click Save-OK; then go to Routine Operations and click Manual Sample ID. Put in sample id by clicking Field and then click OK or press Enter. Click in the following information:
Advia should say Ready to Run. Type strain name like ####F/M##. | ||||||||||||||||||||||
| 10.0 | Notes: n/a |